Recommendation of New Medical Alarms based on Audibility, Identifiability, and Detectability in a Randomized, Simulation-Based Study
Date
2019-08Author
Subject
Metadata
Show full item recordAbstract
<jats:sec> <jats:title>Objectives:</jats:title> <jats:p>Accurate and timely identification of existing audible medical alarms is not adequate in clinical settings. New alarms that are easily heard, quickly identifiable, and discernable from one another are indicated. The “auditory icons” (brief sounds that serve as metaphors for the events they represent) have been proposed as a replacement to the current international standard. The objective was to identify the best performing icons based on audibility and performance in a simulated clinical environment.</jats:p> </jats:sec> <jats:sec> <jats:title>Design:</jats:title> <jats:p>Three sets of icon alarms were designed using empirical methods. Subjects participated in a series of clinical simulation experiments that examined the audibility, identification accuracy, and response time of each of these icon alarms. A statistical model that combined the outcomes was used to rank the alarms in overall efficacy. We constructed the “best” and “worst” performing sets based on this ranking and prospectively validated these sets in a subsequent experiment with a new subject sample.</jats:p> </jats:sec> <jats:sec> <jats:title>Setting:</jats:title> <jats:p>Experiments were conducted in simulated ICU settings at the University of Miami.</jats:p> </jats:sec> <jats:sec> <jats:title>Subjects:</jats:title> <jats:p>Medical trainees were recruited from a convenience sample of nursing students and anesthesia residents at the institution.</jats:p> </jats:sec> <jats:sec> <jats:title>Interventions:</jats:title> <jats:p>In Experiment 1 (formative testing), subjects were exposed to one of the three sets of alarms; identical setting and instruments were used throughout. In Experiment 2 (summative testing), subjects were exposed to one of the two sets of alarms, assembled from the best and worst performing alarms from Experiment 1.</jats:p> </jats:sec> <jats:sec> <jats:title>Measurements and Main Results:</jats:title> <jats:p>For each alarm, we determined the minimum sound level to reach audibility threshold in the presence of background clinical noise, identification accuracy (percentage), and response time (seconds). We enrolled 123 medical trainees and professionals for participation (78 with < 6 yr of training). We identified the best performing icon alarms for each category, which matched or exceeded the other candidate alarms in identification accuracy and response time.</jats:p> </jats:sec> <jats:sec> <jats:title>Conclusions:</jats:title> <jats:p>We propose a set of eight auditory icon alarms that were selected through formative testing and validated through summative testing for adoption by relevant regulatory bodies and medical device manufacturers.</jats:p> </jats:sec>
Collections
Publisher
Place of Publication
Journal
Volume
Issue
Pagination
Recommended, similar items
The following license files are associated with this item:
Related items
Showing items related by title, author, creator and subject.
-
Auditory icon alarms are more accurately and quickly identified than standard melodic alarms in a simulated clinical setting
Mcneer, R; Bodzin Horn, D; Bennett, C; Edworthy, Judy; Dudaryk, R (Lippincott, Williams & WilkinsUnited States, 2018-07)<jats:title>Abstract</jats:title><jats:sec><jats:title>Background</jats:title><jats:p>Current standard audible medical alarms are difficult to learn and distinguish from one another. Auditory icons represent a new type ... -
Acoustic features of auditory medical alarms - an experimental study of alarm volume
Schlesinger,, JJ; Baum, SH; Nash, K; Bruce, M; Ashmead, D; Shotwell, MS; Wallace, MT; Weinger, MB; Edworthy, Judy (Acoustical Society of AmericaUnited States, 2018-06-01)<jats:p>Audible alarms are a ubiquitous feature of all high-paced, high-risk domains such as aviation and nuclear power where operators control complex systems. In such settings, a missed alarm can have disastrous ... -
Can you hear it? Identifying the priority and function of overlapping auditory medical alarms
Parker, Cassie (University of Plymouth, 2021)This paper presents an experimental study that exhibits the differences seen between the recognition of high and moderate priority auditory alarms when played simultaneously. Past research has demonstrated that the current ...