ORCID
- Shakil A Awan: 0000-0002-8543-3328
Document Type
Article
Abstract
We have developed novel impedance standards based on thin-film surface mount device (SMD) resistors. Due to the small dimensions of such resistors, the frequency dependence is very small and can be calculated from quantities that can be either measured or numerically calculated. For nominal resistance values in the 10 kΩ range, as for our application, the DC and AC properties of a single thin-film SMD resistor are not sufficient but can be greatly improved by connecting several thin-film SMD resistors in series. Then the calculated frequency dependence at frequencies up to 2 MHz amounts to only a few parts per million of the DC value, which is about four orders of magnitude smaller than for all conventional calculable AC-DC resistors with a similar nominal DC value. A precision inductance-capacitance-resistance (LCR) meter for frequencies up to 2 MHz, which has a resolution and a reproducibility of a few parts per million but a systematic uncertainty of up to 3000 parts per million, is used to measure two very different SMD-based resistance standards. This enables the verification of the model calculation as well as the calibration of the LCR meter and the correction of its systematic effects, both with a relative uncertainty of a few parts per million in the whole frequency range. This boost in precision enables new applications in this frequency range such as the verification of conventional calculable resistance standards, the calibration of impedance standards, and future measurements of the quantum Hall resistance.
DOI Link
Publication Date
2024-11-25
Publication Title
Measurement Science & Technology
Volume
36
Issue
1
ISSN
0957-0233
Acceptance Date
2024-11-12
Deposit Date
2024-12-17
Funding
This work was not funded or supported by any grant, third party, or company, neither directly nor indirectly.
Additional Links
Keywords
calculable AC-DC resistor, four-terminal-pair impedance standard, precision LCR meter, thin-film SMD resistor
Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 International License.
Recommended Citation
Schurr, J., Judaschke, R., & Awan, S. (2024) 'Resistance standards with very small calculable AC-DC difference at frequencies up to 2 MHz for the calibration of precision LCR meters', Measurement Science & Technology, 36(1). Available at: 10.1088/1361-6501/ad9169