Abstract

The design and performance of a portable, automated flow injection (FI)-based photometric monitor are described. The system is controlled by an in-house microcomputer system that enables the monitor (including a solid state detector) to operate from a 12 V battery supply. The monitor uses the cadmium reduction/diazotization method to analyse for nitrate with a linear range of 0 to 12 mg l-1 and a limit of detection of 0.05 mg l-1 (NO3-N). The hardware and software design, monitor performance and results obtained during unattended operation are presented.

DOI

10.1155/s1463924693000215

Publication Date

1993-01-01

Publication Title

Journal of Automatic Chemistry

Volume

15

Issue

5

Publisher

Hindawi Limited

ISSN

0142-0453

Embargo Period

2024-11-25

First Page

159

Last Page

166